Express Installation and Configuration of TOPSPIN 23
DONE
INDEX
INDEX
a) Type in the NMR ADMINISTRATION PASSWORD
b) Re-enter the NMR ADMINISTRATION PASSWORD
c) Click Next
17. If Data Processing, acquisition and automation or Customized with the Disk-
less package was selected in step 8 a window will appear to inform that the
spectrometer has to be shut down.
a) If the PC is not yet connected to a spectrometer or if the CCU is not booted
1
you can promptly.
• Click Next
b) If your PC is currently connected to a spectrometer and the CCU is booted:
a) Open a Linux shell
b) enter the command:
telnet spect
c) login as root
d) enter the command:
init 5
→ this will shut down the CCU
e) Close the Linux shell
f) Click Next in the window that asked you about shutdown the CCU
Note: The CCU is now either down or switched off and should be turned on again
at the end of this procedure.
→ The installation of T
OPSPIN will start now.
18. After installation is finished a window appears that informs about all installed
programs
• Click Seen
19. Depending if you have installed D
ISKLESS one of the following windows will
appear:
• If you have installed D
ISKLESS a window will appear that asks you to reboot
the PC. Make sure that you can reboot the PC now (check for any unsaved
1. Note that during a first time installation of the TOPSPIN the CCU can not be booted.